8.1.9 Abnormality/Error in the TMS unit (for use with the TMS unit)
If the temperature of the STP pump heating-cooling part is higher than the setting value by 10 ºC (18
ºF) or more due to the occurrence of an abnormality/error in the TMS heater or other fault, the
"FAILURE" LED illuminates and the LCD displays error message "TMS Higher Temp". The heater is
switched OFF and the cooling water unit is operated. The STP pump decelerates and stops. Continue
to operate the cooling water unit while "TMS Higher Temp" is displayed after the pump stops.
If the temperature of the STP pump heating-cooling part is lower than the setting value by 10 ºC (18 ºF)
or more due to the occurrence of an abnormality/error in the TMS heater or other fault, the "FAILURE"
LED illuminates and the LCD displays the error message "TMS Lower Temp". The heater and the
cooling water valve are switched OFF and the STP pump decelerates and stops.
Note: "TMS Lower Temp" is not detected for a certain time when the STP pump starts.
When an abnormality/error occurs in the TMS connection cable, the STP pump decelerates and stops.
The heater and the cooling water unit are switched OFF.
The "FAILURE" LED illuminates and the LCD displays the error message "T.Cable Disconnected".
8.1.10 Failure of the air-cooling fan
When the air-cooling fan in the SCU-801 locks and stops, the "FAILURE" LED illuminates and the LCD
displays error message "Driver Failure" and "Driver Fan Failure". The STP pump decelerates and
stops.