SCU-801 Control Unit for Turbomolecular Pump
MT-95E-001-A
Page 75
5
SERIAL COMMUNICATION PROTOCOL
However, when the MSB (most significant bit) is always 0 when data length is 7 bits, LRC is set to
6C
hex
.
5.3.7 Error control
1. Transmit the transmission frame repeatedly from the PC when the SIM transmits "Nak" (parity
check error). When the SIM receives "Nak" from the PC, the transmission frame is transmitted
again. This operation is repeated up to 5 times.
2. The SIM transmits "Ack" or "Nak" to the PC after the completion of communication command
reception. When the PC cannot receive "Ack" or "NaK" after 2 second, retransmit the transmission
frame from the PC.
When these communication statuses occur repeatedly, display to an error message or start the error
routine on the PC.