EasyManua.ls Logo

Edwards SCU-801

Edwards SCU-801
149 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
SCU-801 Control Unit for Turbomolecular Pump
MT-95E-001-A
Page 54
4
OPERATION
Pin
Description
Pins for outputting the STP pump REMOTE SELECTION state
signal.
When the MANUAL/REMOTE changeover switch on the front
panel is set to ON (remote side), these pins are closed
("REMOTE" operation).
Pins for outputting the POWER ON state signal.
These pins are closed when magnetic bearing functions and
the rotor levitates. This output is opened at a power failure.
Pins for outputting the ACCELERATION state signal.
These pins are closed when the STP pump is in acceleration.
Pins for outputting the NORMAL OPERATION state signal.
When the STP pump is in rated operation, the pins between
(11)-(30) are closed, and the pins between (12)-(30) are
opened.
Pins for outputting the BRAKE state signal.
These pins are closed when the STP pump is in brake.
Pins for outputting the ALARM signal.
When an abnormality/error is detected under the POWER ON
state, the pins between (14)-(33) are closed, and the pins
between (15)-(33) are opened.
Pins for the TMS unit rated state signal output.
When the actual temperature of the TMS unit is within the
setting temperature range ±10°C, the pins between (16)-(17)
are closed and the pins between (35)-(17) are opened.
Pins for the WARNING signal output. When the WARNING
message is displayed on the LCD, the pins between (24)-(6)
are closed and the pins between (25)-(6) are opened.
Pins for the emergency vent valve operation signal
(optional accessory). These pins are closed while this valve
functions.
Table 10 - REMOTE X7 output signal pins
L.VALVE
N.O OUT
(25)
(24)
(6)
WARNING OUT
COM
N.C
N.O

Table of Contents

Related product manuals