EasyManua.ls Logo

Edwards SCU-801

Edwards SCU-801
149 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
SCU-801 Control Unit for Turbomolecular Pump
MT-95E-001-A
Page 86
5
SERIAL COMMUNICATION PROTOCOL
5.4.2 ReadMeas
Function: Reads the measured rotational speed.
Transmission frame:
PC->SIM
Stx
0
0
1
?
D
Etx
LRC
SIM->PC
Ack
PC->SIM
Ack
SIM->PC
Stx
0
0
1
Sp
D
Parameter 1 to 2
Etx
LRC
Parameter
Item
Data format
Remarks
1
[System reservation]
56-bits hexadecimal coded
ASCII
2
Measured rotational speed
(Unit: Hz)
16-bits hexadecimal coded
ASCII
Example:
Measured rotational speed: 02DC
hex
= 732 Hz = 43,920 rpm
Parameter
1
2
ASCII
*1
*1
*1
*1
*1
*1
*1
*1
*1
*1
*1
*1
*1
*1
0
2
D
C
HEX
30
32
44
43
*1 System reservation
5.4.3 Command
Function: Sends the pump operation commands START, STOP and RESET. These commands
function in the same way as when each switch of the SCU-801 is pressed. They are valid when
MANUAL/REMOTE changeover switch on the SCU-801 is "ON". Refer to Section 3.11, "Parameter set
mode".
Transmission frame:
PC->SIM
Stx
0
0
1
Sp
E
Parameter 1
Etx
LRC
SIM->PC
Ack
PC->SIM
Ack
SIM->PC
Stx
0
0
1
#
Etx
LRC

Table of Contents

Related product manuals