exposure time for 20 minutes.
Important: Temperature should not exceed 121°C.
Important: Exposure time should not exceed 20 minutes.
11.12 Visually Inspecting the Mirror in the Scanner Tip
Before scanning, verify that the mirror inside the end of the tip does not have any
smudges or scratches on it. This can affect image quality. Replace the tip as necessary.
11.13 Visually Inspecting the Scanner Tips for Damage
Visually inspect the scanner tips for signs of deterioration. Discard if any damage
is noted.