Electrostatic
discharge
(ESD)IEC
61000-4-2
±8 kV
contact
±15 kV air
±8 kV
contact
±15 kV air
Floors should be wood, concrete,
or ceramic tile. If floors are covered with
synthetic material, the relative humidity
should be at least 30%.
Power
frequency
(50/60 Hz)
magnetic
field IEC
61000-4-8
30 A/m
30 A/m
Power frequency magnetic fields
should be at levels characteristic of a
typical location in a typical commercial or
hospital environment.
Radiated RF
IEC 61000-
4-3
3 V/m
80MHz
2.7GHz
3 V/m
80MHz
2.7GHz
Environment of a professional
healthcare facility. WARNING: Portable RF
communications equipment
(including peripherals such as antenna
cables and external antennas) should be
used no closer than 30 cm (12 inches) to
any part of the HELIOS 600 including
cables specified by the manufacturer.
Otherwise, degradation of the
performance of this equipment could
result.
13.2.2 For the immunity to proximity fields from RF wireless communications
equipment, the HELIOS 600 is compliant with the test levels specified below,
according to IEC 60601-1-2 standard. The customer or user of the HELIOS 600
should assure that it is used in such an environment.
Test
Frequency
(MHz)
Band (MHz)
Immunity Test Levels
385
380 390
Pulse modulation 18Hz, 27V/m
450
430 470
FM, ±5 kHz deviation, 1 kHz sine, 28V/m
710
704 787
Pulse modulation 217Hz, 9V/m
745
780
810
800 960
Pulse modulation 18Hz, 28V/m