DISPLAY ERROR MESSAGES
When an error occurs the error number will be displayed. The probable cause of each error
is
listed below.
OPERATOR ERRORS
The following error messages indicate an operator error.
01 Illegal Key Sequence.
02 A resolution number was not entered.
03 A band number was not entered; or the number entered was
too large.
04 No power reading in current band.
05 Frequency limit high
>
18.5 GHz, 27 GHz (548A).
06 (Freq Limit HI)
-
(Freq Limit Lo)
<
Min. (100 MHz) difference.
07 Frequency Limit Low
<
.95 GHz (545A1548A).
08
09 Illegal test mode key sequence.
10 Illegal DAC key sequence.
1 1
Illegal Multiplier key sequence.
12
13 Option not installed.
COUNTER ERRORS
The following error messages indicate a malfunction within the counter.
31 Check sum error Section 3 PROM
DO000
-
DFFF A105, U13
32 Check sum error Section 2 PROM
E000- EFFF A105, U17
33 Check sum error Section 1 PROM
E000- FFFF A105, U15
34 Check sum error Power Meter PROM 4000
-
47 F F A107, U20
35 Check sum error Band 4 PROM
COO0
-
C7 FF A105, U14
36 Check sum error
GPlB or BCDIRemote PROM C800
-
CFFF A105, U16