21. Push the auto cleaner up into the oven and hold it there.
22. Carefully retighten the two previously loosened screws completely.
23. Release the auto cleaner.
24. Connect the two compressed air lines of the auto cleaner to the control assembly (blue to
blue, black to black).
25. Reconnect the analyser to the compressed air and power supply.
26. Set a pressure of 1 bar on the control assembly.
27. Fit the service cover supplied.
28. Connect the vacuum cleaner to the adapter. Secure the hose to the adapter with the hose
clip.
29. Insert the adapter and secure it turning.
30. Plug the vacuum cleaner into the socket on the back of the analyser. The auto cleaner is
installed and aligned.