15. Select an application and work through the application-specific paramaters for the auto
cleaner. You can set the frequency of automatic cleaning in the application parameters
(cleaning interval):
1 =after every analysis; 10 = after every 10th analysis.
16. You can similarly set the cleaning duration in seconds.
17. Save the changes.
18. Fit the top right service cover using the cover supplied.
The device has been aligned and put into operation.
The device is ready for analysis after approx. 30 to 45 minutes once the cells have been
thermostated (base lines stable) and the catalysis oven has reached operating temperature.