Risk for eye injuries
Chemicals
− When chemicals are replaced, small particles of the chemicals can float in
the air and burn the eyes.
• Please always wear protective goggles when working with
chemicals.
• Please observe the safety data sheets in terms of the chemicals
used.
The procedure described in this section applies to the filters on the front side of the analyser
behind the device door. The reagent tubes of these filters must also be removed in the same way
for filling or replacement.
In order to remove a reagent tube on the front of the analyser, please proceed as follows:
1. Turn the main switch of the analyser to switch position 1.
2. In order to release the pressure, press the “Reagent Change” button on the “Status
Analyser” page in the ELEMENTS software.
3. Slide the reagent tube upwards.
4. Swivel the reagent tube slightly to the side or forwards.
5. Pull the reagent tube downwards and out. If the reagent tube is difficult to be moved, turn it
slightly while pulling it out.
The reagent tube on the front side of the analyser has been removed.
The following figure shows an example of how to remove a reagent tube on the front side of the
analyser: