PARTNER
EMC
4 Perform a Test
For 3-phase AC, connect L1, L2, L3, N and PE
For DC (floating), connect DC+ to L1 and DC- to N
For DC (when DC- = PE), connect DC+ to L1 and DC- to PE
The fused input is only for AC. For DC, the dedicated unfused DC input must be used.
4.4 Slow DOW Magnetic Field Test
1
2
2a
3
1 Generator
Select the magnetic field test and the correct antenna as coupling device.
A missing coupling device in the parameter list can be added in the generator main menu.
2 Antenna
The DOW3000 can be used either with the MF1000-1 or MF1000-2. Connect the antenna with a
2m MC-cable to the direct output of the generator.
The antenna must be configured with all bridges in the lower position as shown in the
picture.
The mains socket and the the range selection switch are not used for this test.
3 EUT
Place the EUT in the center of the antenna
Page 21/55