A.2.5 Fieldbus combinations
Table A-1: Fieldbus Combination Matrix (Non-SIL)
Primary Fieldbus options
TRL2 RS485 Enraf Analog out
passive (non-
IS)
Analog In
passive (non-
IS)
Secondary Fieldbus
options
Code R 4 E B 7
TRL2 R Yes Yes No No No
Enraf E Yes Yes No No No
WirelessHART
®
W Yes Yes Yes Yes Yes
L&J Tankway 1500 XL/MCG
2000
L Yes Yes No No No
Varec Mark/Space GT
1800/1900
V Yes Yes No No No
Whessoe WM 550/660
(digital current loop)
H Yes Yes No No No
GPE 31422/31423 (digital
current loop)
G Yes Yes No No No
Sakura MDP/V1 U Yes Yes No No No
Tokyo Keiso T Yes Yes No No No
Analog out active (IS) C Yes Yes Yes No No
Analog out active (non-IS) A Yes Yes Yes No No
Analog out passive (IS) D Yes Yes Yes No No
Analog out passive (non-IS) B Yes Yes Yes No No
Analog in active (IS) 8 Yes Yes Yes No No
Analog in active (non-IS) 6 Yes Yes Yes No No
Analog in passive (IS) 9 Yes Yes Yes No No
Analog in passive (non-IS) 7 Yes Yes Yes No No
None 0 Yes Yes Yes No No
Ready for upgrade F Yes Yes Yes No No
Yes = Primary Fieldbus and Secondary Fieldbus can be combined
No = Combination not possible
Specifications and reference data Reference Manual
December 2019 00809-0100-2410
120 Reference Manual