Figure A-1: Maximum Upper Product Thickness for the Flexible Single Lead Probe
1
3 5
7
9
11
0
16 (5)
33 (10)
49 (15)
66 (20)
82 (25)
98 (30)
115 (35)
A
B
C
A. Maximum upper product thickness, ft. (m)
B. Upper product dielectric constant
C. Lower product dielectric constant
Figure A-2: Maximum Upper Product Thickness for the Flexible Twin Lead Probe
1
3 5
7
9
11
0
16 (5)
33 (10)
49 (15)
66 (20)
82 (25)
98 (30)
115 (35)
A
B
C
A. Maximum upper product thickness, ft. (m)
B. Upper product dielectric constant
C. Lower product dielectric constant
A.1.4
Accuracy over measuring range
The measuring range depends on probe type, dielectric constant of the product and
installation environment, and is limited by the Blind Zones at the very top and bottom of
the probe. In the Blind Zones, the accuracy exceeds ±1.18 in. (30 mm), and measurements
may not be possible. Measurements close to the Blind Zones will have reduced accuracy.
Specifications and reference data Reference Manual
April 2019 00809-0100-4530
276 Rosemount 5300 Level Transmitter