Ui Ii Pi Ci Li
Entity parameters
Fieldbus
30 V 300 mA 1.5 W 4.95 nF 0 mH
B.10.3 IC China FISCO
Certificate
GYJ16.1095X
Standards
GB 3836.1/2/4/20-2000, GB 12476.4/5-2013, GB 12476.1-2010
Markings
Ex ia IIC T4 Ga (-50 °C ≤ Ta ≤ +60 °C)
Ex ia/ib IIC T4 Ga/Gb (-50 °C ≤ Ta ≤ +60 °C)
Ex iaD 20 T69 (-50 °C ≤ Ta ≤ +60 °C)
Ex iaD/ibD 20/21 T69 °C (-50 °C ≤ Ta ≤ +60 °C)
Special Conditions for Safe Use (X):
1. See certificate for Specific Conditions.
Ui Ii Pi Ci Li
FISCO parameters 17.5 V 380 mA 5.32 W 4.95 nF <0.001 mH
B.10.4 N3 China Type N
Certificate
GYJ18.1331X
Standards
GB 3836.1-2010, GB 3836.4-2010, GB 3836.8-2014
Markings
Ex nA ic IIC T4 Gc (-50 °C ≤ Ta ≤ +60 °C/+70 °C)
Ex ic IIC T4 Gc (-50 °C ≤ Ta ≤ +60 °C/+70 °C)
Special Conditions for Safe Use (X):
1. See certificate for Specific Conditions.
Ui Ii Pi Ci Li
Safety parameters HART 42.4 V 23 mA 1 W 7.25 nF Negligible
Safety parameters
Fieldbus
32 V 21 mA 0.7 W 4.95 nF Negligible
B.11 Technical Regulations Customs Union (EAC)
B.11.1 EM Technical Regulations Customs Union (EAC)
Flameproof
Certificate
RU C-SE.AA87.B.00802