5. If the transmitter head is not mounted directly, attach the protection plug to the
process seal to protect the exposed parts from dust and water. If a protection plug
is not available, then cover the process seal with a plastic bag.
6. Rotate the new transmitter head so the device display faces the desired direction.
7. Tighten the nut. Max torque is 30 ft-lb (40 Nm).
Postrequisites
Configure the transmitter.
Related information
Configuration
7.13 Changing a probe
7.13.1 Probe and firmware compatibility
Transmitter heads with a firmware version earlier than 1.A4 (manufacturing date before
2008-06-18) are not compatible with HP/HTHP/C probes marked with R2.
Transmitter heads with a firmware version 1.A4 or later are compatible with HP/HTHP
probes without the R2 marking when the Trim Near Zone function is performed, as
illustrated below.
Only probes with both R3 and VC marking are compatible with the Dynamic Vapor
Compensation function. See Check if Dynamic Vapor Compensation function is supported
to determine if the Dynamic Vapor Compensation function is supported by the
transmitter head.
Service and troubleshooting Reference Manual
April 2019 00809-0100-4530
238 Rosemount 5300 Level Transmitter