B.9.3 IB INMETRO FISCO
Certificate
UL-BR 17.0188X
Standards
ABNT NBR IEC 60079-0:2013, ABNT NBR IEC 60079-11:2013, ABNT NBR IEC
60079-26:2016, ABNT NBR IEC 60079-31:2014
Markings
Ex ia IIC T4 Ga (- 55 °C ≤ T
amb
≤ +60 °C)
Ex ia/ib IIC T4 Ga/Gb (- 55 °C ≤ T
amb
≤ +60 °C)
Ex ia IIIC T69 °C Da (- 50 °C ≤ T
amb
≤ +60 °C)
Ex ia/ib IIIC T69 °C Da/Db (- 50 °C ≤ T
amb
≤ +60 °C)
Special Conditions for Safe Use (X):
1. See certificate for Specific Conditions.
Ui Ii Pi Ci Li
FISCO parameters 17.5 V
dc
380 mA 5.32 W 4.95 nF <1 µH
B.10 China
B.10.1 E3 China Flameproof
Certificate
GYJ16.1095X
Standards
GB 3836.1/2/4/20-2010, GB 12476.1/5-2013, GB 12476.4-2010
Markings
Ex d ia IIC T4 Ga/Gb (-40 °C ≤ Ta ≤ +60 °C/+70 °C)
Ex tD A20 IP 66/67 T69 °C /T79 °C (-40 °C ≤ Ta ≤ +60 °C/+70 °C)
Special Conditions for Safe Use (X):
1. See certificate for Specific Conditions.
B.10.2
I3 China Intrinsic Safety
Certificate
GYJ16.1095X
Standards
GB 3836.1/2/4/20-2010, GB 12476.1/5-2013, GB 12476.4-2010
Markings
Ex ia IIC T4 Ga (-50 °C ≤ Ta ≤ +60 °C/+70 °C)
Ex iaD 20 T69 °C /T79 °C (-50 °C ≤ Ta ≤ +60 °C/+70 °C)
Ex iaD/ibD 20/21 T69 °C (-50 °C ≤ Ta ≤ +60 °C)
Special Conditions for Safe Use (X):
1. See certificate for Specific Conditions.
Ui Ii Pi Ci Li
Entity parameters HART 30 V 130 mA 1 W 7.26 nF 0 mH
Product certifications Reference Manual
April 2019 00809-0100-4530
368 Rosemount 5300 Level Transmitter