Procedure
1. Select Setup Tank.
2. Select the Environment tab.
3. Select the Upper Product Dielectric Constant Calculator button.
C.2.4 Volume configuration
For volume calculations, you can choose one of the standard tank shapes or the strapping
option. Choose None if volume calculation is not used. For the standard tanks, a volume
offset parameter can be specified which can be used for a non-zero volume that
corresponds to the zero level. This may be useful, for example, if the user wants to include
the product volume below the zero level.
Tank type
You can choose one of the following options:
Strap table
Vertical cylinder
Horizontal cylinder
Vertical bullet
Horizontal bullet
Sphere
None
Strapping table
Use a strapping table if a standard tank type does not provide sufficient accuracy. Use
most of the strapping points in regions where the tank shape is non-linear. A maximum of
20 points can be added to the strapping table.
If tank type Strapping Table was chosen, enter how many entries you will use and the
actual level and volume points. The strapping points must be entered such that the first
point corresponds to the lowest level, and the last point corresponds to the topmost level
of the tank.
Configuration parameters Reference Manual
April 2019 00809-0100-4530
390 Rosemount 5300 Level Transmitter