Figure 3-9: Insulated Chamber
See Insulated tanks for more information.
3.2.7
Installation in non-metallic tanks and open-air
applications
Avoid major sources of electrical disturbance in proximity of the installation, e.g. electrical
motors, stirrers, servo mechanisms.
Figure 3-10: Avoid Electromagnetic Disturbances
For clean liquids, use a coaxial probe to reduce effect of potential electrical disturbances.
Figure 3-11: Coaxial Probe in an Open-Air Application
Mechanical installation Reference Manual
April 2019 00809-0100-4530
40 Rosemount 5300 Level Transmitter