Procedure
1. In Rosemount Radar Master, select Setup → Tank.
2. Select the Environment tab.
3. Select the Advanced button for more options.
4. In the Vapor Dielectric Constant box, the desired value. You may also use the Vapor
Dielectric Calculator or the Vapor Dielectric Chart to find the correct value.
Figure D-5: Dielectric Constants Can Be Adjusted in the Tank Environment
Window
D.5.2 Adjust the lower product dielectric range
If the dielectric constant of the lower product is significantly smaller than the dielectric
constant of water, you may need to adjust the Lower Product Dielectric Range.
Prerequisites
Special adjustments can also be made by trimming the appropriate amplitude thresholds.
See Example 2: Interface peak not found for further information.
Procedure
1. In Rosemount Radar Master, select Setup → Tank.
2. Select the Environment tab.
3. Select the Advanced button for more options.
4. In the Lower Product Dielectric Range list, select the desired range.
Advanced configuration Reference Manual
April 2019 00809-0100-4530
404 Rosemount 5300 Level Transmitter