7. Electronics Failure / Output Board
8. Software Incompatibility
ErrorFAILED_MASK
This parameter will mask any of the failed conditions listed in FAILED_ENABLED. A bit on
means that the condition is masked out from alarming and will not be reported.
FAILED_PRI
Designates the alerting priority of the FAILED_ALM, see Alarm priority. The default is 0 and
the recommended values are between 8 and 15.
FAILED_ACTIVE
This parameter displays which of the alarms is active. Only the alarm with the highest
priority will be displayed. This priority is not the same as the FAILED_PRI parameter
described above. This priority is hard coded within the device and is not user configurable.
FAILED_ALM
Alarm indicating a failure within a device which makes the device non-operational.
I.3.2
MAINT_ALARMS
A maintenance alarm indicates that the device or some part of the device needs
maintenance soon. If the condition is ignored, the device will eventually fail. There are five
parameters associated with MAINT_ALARMS, they are described below.
MAINT_ENABLED
The MAINT_ENABLED parameter contains a list of conditions indicating that the device or
some part of the device needs maintenance soon.
Below is a list of the conditions:
1. Configuration Error
2. Configuration Warning
3. Simulation Mode
4. Volume / Temperature Measurement Warning
5. Vapor Compensation Warning
6. Probe Contamination
MAINT_MASK
The MAINT_MASK parameter will mask any of the failed conditions listed in
MAINT_ENABLED. A bit on means that the condition is masked out from alarming and will
not be reported.
MAINT_PRI
MAINT_PRI designates the alarming priority of the MAINT_ALM, see Process alarms. The
default is 0 and the recommended values are 3 to 7.