Section 5: Basic Configuration/Start-up
November 2014
Basic Configuration/Start-up
5.4.9 Using the Setup functions
Use the Setup function if you are already familiar with the configuration process for the
Rosemount 5400 Series transmitter or for changes to the current settings:
Figure 5-10. Setup Functions in RRM
1. Start the RRM software.
2. In the RRM workspace, select the
appropriate icon for the configuration
of transmitter parameters:
Wizard: the Wizard is a tool that
guides you through the basic
configuration procedure of a
Rosemount 5400 Series
transmitter
General: configuration of general
settings, such as measurement
units and communication
parameters. This window also lets
you configure which LCD
variables to be displayed.
Tank: configuration of Tank
Geometry, Tank Environment,
and Volume
Echo Curve: disturbance echo
handling
Advanced: advanced
configuration
General
Echo Curve
Advanced
Wizard
Tank geometry,
Environment,
Volume