126
Reference Manual
00809-0100-4026, Rev HA
Section 6: Operation
November 2014
Operation
6.3 LCD display error messages
Figure 6-7. The Rosemount 5400 Series Display Panel Displaying an Error Message
Table 6-2. Error Messages Displayed on the Rosemount 5400 Series Display Panel
For more information on errors, see “Error Messages” on page 154.
Error message Description
RAM FAIL
An error in the gauge data memory (RAM) has been detected during the
startup tests.
NOTE: this resets the gauge automatically.
FPROM FAIL
An error in the gauge program memory (FPROM) has been detected
during the startup tests.
NOTE: this resets the gauge automatically.
HREG FAIL
An error in the transmitter configuration memory (EEPROM) has been
detected. The error is either a checksum error that can be solved by
loading the default database or a hardware error.
NOTE: the default values are used until the problem is solved.
OMEM FAIL Other memory failure.
MWM FAIL An error in the microwave module.
DPLY FAIL An error in the LCD display.
MODEM FAIL Modem hardware failure.
AOUT FAIL An error in the Analog Out Module.
OHW FAIL An unspecified hardware error has been detected.
ITEMP FAIL An error in the internal temperature measurement.
MEAS FAIL A serious measurement error has been detected.
CONFIG FAIL
At least one configuration parameter is outside the allowed range.
NOTE: the default values are used until the problem is solved.
SW FAIL An error has been detected in the transmitter software.
Error message