221
Reference Manual
00809-0100-4026, Rev HA
Appendix B: Product certifications
November 2014
Product Certifications
B.5.2 Canadian Standards Association (CSA) Approvals
When bearing the “Dual Seal” marking, this product meets the Dual Seal Requirements of
ANSI/ISA 12.27.01-2003.
Cert. No.: 1514653
E6 Explosion-proof with internal Intrinsically safe circuits [Exia] Class I, Div. 1, Groups B, C,
and D;
Temp. Code T4.
Class II, Div. 1 and 2, Groups E, F, and G;
Class III, Div. 1
Ambient temperature limits -50 °C to +70 °C
(1)
.
Approval valid for HART, F
OUNDATION fieldbus, and Modbus options.
I6, IF Intrinsically safe Exia.
Class I, Div. 1, Groups A, B, C, and D.
Temp. Code T4.
4-20 mA/HART model: U
i
= 30 Vdc, I
i
= 130 mA, P
i
= 1.0 W, C
i
= 7.26 nF, L
i
= 0 H.
F
OUNDATION fieldbus model: U
i
= 30 Vdc, I
i
= 300 mA, P
i
= 1.3 W, C
i
= 0 nF, L
i
= 0 H.
FISCO model: U
i
= 17.5 Vdc, I
i
= 380 mA, P
i
= 5.32 W, C
i
= L
i
= 0.
Installation drawing: 9150079-906.
Ambient temperature limits: -50
°C to +70 °C
(1)
.
Approval valid for HART, F
OUNDATION fieldbus, and FISCO options.
(1) +60 °C with FOUNDATION fieldbus or FISCO option.