244
Reference Manual
00809-0100-4026, Rev HA
Appendix C: Advanced configuration
November 2014
Advanced Configuration
Empty tank detection area
If the signal from the product surface is lost within the region given by the parameter Empty
Tank Detection Area, the tank is considered empty and the product level is presented as zero.
If the surface is lost above the Empty Tank Detection Area, the transmitter starts searching the
entire tank for the surface.
The Empty Tank Detection Area can be increased if the surface is lost outside the Empty Tank
Detection Area in a non-critical region of the tank.
1. Disable Use Automatic Empty Tank Handling Settings.
2. Type the desired value in the Empty Tank Detection Area input field.
Amplitude
Distance
Echo from
tank bottom
Surface echo