TROUBLESHOOTING
Symptom Solution
No or low irrigation ow
Check if the instrument is blocked. Blow compressed air through the instrument to unblock it
Use another handpiece to check if the handpiece is blocked
If the obstruction cannot be removed, the handpiece must be sent to an approved EMS repair center
Please refer to the operation instructions of the Piezon unit for more details
No ultrasound oscillations
Please refer to the operation instructions of the Piezon unit
Decreasing or reduced
performance
Follow the instructions found in the operation instructions of the Piezon unit
Make sure that the instrument is properly positioned, tighten it again with the CombiTorque
Check the instrument for wear and replace it if needed
Try to use the handpiece with another instrument
Send the handpiece and the instrument to an approved EMS repair center
Breakage of a le or instrument,
possibly inside the cavity or the
root canal
Make sure that every broken part is located
Compare total length of the broken parts with the length of a new le or instrument to make sure that all broken parts
are present
Try to remove broken parts inside the root canal by activating maximum irrigation (no ultrasound) on a le to attempt
to force the debris out
Refer to instructions and use the RT2, RT3 and RT4 instruments that are specially designed for this application
Diamond-coated instruments
become less effective
Visually check the condition of the diamond coating and replace the instrument if needed
FB-439_3_Piezon instruments.indd 112 28.10.2015 14:49:50