EasyManua.ls Logo

Endress+Hauser Analytik Jena Biometra TRIO Series - Error-Logfile; Extended Self-Test

Endress+Hauser Analytik Jena Biometra TRIO Series
110 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Biometra TRIO Tools
Version 03_19 89
11.3.4 Error-Logfile
The Biometra TRIO Thermal Cycler records all errors by time and date. The latest error
is shown in position in the Error Log table. Additionally the error code and error mes-
sage are listed.
Figure 70 Biometra TRIO Error-Logfile screen
11.4 Extended Self-Test
NOTICE
The Extended Self-Test must be performed under the following conditions to achieve
reproducible and meaningful results:
Ambient temperature in the range 20 °C to 25 °C
Mains voltage depending on the voltage type of the device: 100 V/115 V/230,
±10 % each
The device must be cooled to the ambient temperature of 20 °C to 25 °C after the
last application.
During the Extended Self-Test the Biometra TRIO Thermal Cycler checks several func-
tions and components:
Test
Function
COOLER
Incubates the sample block to 4°C and checks, if the sample block
reaches 4°C and if the low temperature is hold for a longer time.
THERMAL TRACKING
Checks the synchronism of the control circuits and if they work coordi-
nated together.
HEATING AND
COOLING RATE
Checks, if the specified average heating and cooling rate is reached.
REFRIGERATION
The cooling sink and fans are checked to work properly together.

Table of Contents

Related product manuals