Biometra TRIO Tools
Version 03_19 91
Figure 72 Biometra TRIO Extended Self-Test screen
During the test the software provides a progress overview. If a test is already passed
the result “Failed” or “Passed” is displayed, for the currently performed test “Running”
is shown and all other tests are not finished yet.
Note: The Extended Self-Test will take approximately 30 minutes to finish and the
instrument is blocked, no other program can be started while the test is running. To
stop the Extended Self-Test press:
STOP EXTENDED SELF-TEST
If the test is stopped, no corresponding log file will be created.
Figure 73 Biometra TRIO Extended Self-Test progress overview screen
The results of the Extended Self-Test are summarized and stored by the instrument
and can be viewed by the user (see chapter "Extended Self-Test-Logfile" p. 88).