Commissioning Gammapilot FMG50
54 Endress+Hauser
0%
10000
100%
I
N
L (%)
A0040241
 18 Example of a linearization curve for level measurements (consisting of 6 value pairs)
L Level
I
N
Normalized pulse rate
The type of linearization has already been selected in the "Measurement settings" section
The behavior of the "linear" type of linearization is identical to the "point level
calibration" operating mode.
1. Selection: start with full calibration or start with empty calibration
Start calibration -> the calibration can be stopped once the pulse rate has
stabilized.
A0042168