2. Empty calibration level: the radiation is switched on and the beam path is
completely free.
If these conditions are met, empty calibration can be started.
A0042169
The empty calibration can be performed by pressing the "Start empty calibration"
button. The measurement then starts automatically and continues, at the very
maximum, for as long as has been configured for the calibration time. However,
the process can also be stopped manually by pressing the "Stop calibration"
button.
The calibration stops automatically as soon as a million pulses have been
totalized.
Alternatively, the empty calibration can also be entered directly.
For the "Next" button in the Wizard to be enabled, the value must, however, be
changed from the start value, at least temporarily.
3. Full calibration level: the radiation is switched on and the beam path is completely
covered by medium.
If these conditions are met, the calibration can be started.
A0042170
The full calibration can be performed by pressing the "Start full calibration"
button. The measurement then starts automatically and continues, at the very
maximum, for as long as has been configured for the calibration time. However,
the process can also be stopped manually by pressing the "Stop calibration"
button.
The calibration stops automatically as soon as a million pulses have been
totalized.
Alternatively, the full calibration can also be entered directly.
For the "Next" button in the Wizard to be enabled, the value must, however, be
changed from the start value, at least temporarily.
TIP: if the vessel cannot be filled appropriately, the full calibration can also be
performed with the radiation switched off. This is a way of simulating a
completely covered radiation path. In this case, full calibration is identical to
background calibration and 0 cnt/s is typically displayed.