Gammapilot FMG50 Commissioning
Endress+Hauser 69
2. Selection: start with a high concentration of the radiating medium or start with a
low concentration of the radiating medium
Start calibration -> the calibration can be stopped once the pulse rate has
stabilized.
A0042222
3. Calibration with high concentration
Press the "Calibration conc. self-rad. high" button
4. Calibration with low concentration
Press the "Calibration conc. self-rad. low" button
5. The measurement then starts automatically and continues, at the very maximum, for
as long as has been configured for the calibration time.
However, the process can also be stopped manually by pressing the "Stop
calibration" button.
The calibration stops automatically as soon as a million pulses have been
totalized.
6. Entry for each calibration point: enter the concentration of the medium in the
"Calibration conc. self-rad. high" and "Calibration conc. self-rad. low" field
This establishes the reference between the determined pulse rate and the
concentration of the radiating medium.
TIP: take a sample of the medium during the integration and then determine the
concentration (e.g. in the laboratory)
7. If a customized table has been selected for the linearization, the following input
screen appears:
A0042223