Commissioning Gammapilot FMG50
68 Endress+Hauser
5. The calibration has been performed successfully.
A0042220
6. The settings for the current output are then made in the "Output settings" step
Concentration of radiating media
For concentration measurement in radiating media (e.g.: K40), the Gammapilot FMG50
requires at least two other calibration points in addition to the background calibration:
• Pulse rate at high concentration of radiating medium
• Pulse rate at low concentration of radiating medium
The linearization defines the correlation between the measured pulse rate and the
concentration of the radiating medium (0 to 100%).
The Gammapilot FMG50 makes a variety of linearization modes available:
• Linear assignment of the pulse rate to the concentration
• Entry of any linearization table adapted to the specific application.
• The linearization table consists of up to 32 "normalized pulse rate : concentration"
value pairs
• The linearization table must be monotonic increasing, i.e. a higher concentration must
always be paired with a higher pulse rate.
A0042221
20 Example of a linearization curve for measurements of the concentration of radiating media
C Concentration of radiating media
I
N
Normalized pulse rate
1. Selection of the type of linearization (already selected in the "Measurement settings"
section)