Levelflex FMP51, FMP52, FMP54 Description of device parameters (operating menu)
Endress+Hauser 119
DC value
Navigation
Setup ® DC value
Conditions
Only visible for devices with "Interface measurement" application package (Product structure:
Feature 540 "Application package", option EB "Interface measurement")
Description
Defines the relative dielectric constant e
r
of the
upper phase (DC
1
).
DC
1
DC
1
A0013181
Input range
0 to 20
Factory setting
1.9
Additional information
Dielectric constants of important media commonly used in the industry are summarized in the
document SD106F, which can be downloaded from the Endress+Hauser web page
(www.endress.com).
Medium group
Navigation
Setup ® Medium group
Condition
Only visible for "Operating mode" = "Level".
Description
Defines the medium group of the measured product.
Selection
Water bsed DC > 4
Others DC ³ 1.9
Factory setting
Others DC ³ 1.9
Empty calibration
Navigation
Setup ® Empty calibration