Description of device parameters (operating menu) Levelflex FMP51, FMP52, FMP54
126 Endress+Hauser
Description
Use this function to enable write-protected parameters via local operation or an operating tool.
For local operation, the customer-specific access code defined in the Define access code
parameter is entered (® ä 126). If an incorrect access code is entered, the user retains his
current access authorization.
The write protection affects all parameters marked with the symbol in the document. On the
local display, the symbol in front of a parameter indicates that the parameter is write-protected.
If no key is pressed for 10 minutes, or the user goes from the navigation and editing mode back to
the measured value display mode, the device automatically locks the write-protected parameters
after another 60 s .
Please contact your Endress+Hauser Sales Center if you lose your access code
Input range
0 to 9 999
Define access code
Navigation
Setup ® Advanced setup ® Define access code
Description
Use this function to restrict write-access to parameters to protect the configuration of the device
against unintentional changes via local operation. A user-specific access code is specified for this
purpose.
The write protection affects all parameters marked with the symbol in the document. On the
local display, the symbol in front of a parameter indicates that the parameter is write-protected.
Once the access code has been defined, write-protected parameters can only be modified if
the access code is entered in the Enter access code parameter (® ä 125).
Changing the access code
•
Enter the current access code in the Enter access code parameter and confirm.
• Define the new access code.
Please contact your Endress+Hauser Sales Center if you lose your access code
Input range
0 to 9 999
Factory setting
0
Additional information
User entry
A message is displayed if the access code is not in the input range.
Factory setting
If the factory setting is not changed or 0 is defined as the access code, the parameters are not
write-protected and the configuration data of the device can then always be modified. The user is
logged on in the role of Maintenance .
Device tag
Navigation
Setup ® Advanced Setup ® Device tag