Levelflex FMP51, FMP52, FMP54 Description of device parameters (operating menu)
Endress+Hauser 133
Factory setting
80 (Dielelectric constant of water at 20 °C / 68 °F)
Additional information
Dielectric constants of important media commonly used in the industry are summarized in
the document SD106F, which can be downloaded from the Endress+Hauser web page
(www.endress.com).
Level unit
Navigation
Setup ® Advanced setup ® Interface ® Level unit
Description
Defines the level unit
Options
• %
•
m
• mm
• ft
• in
Factory setting
%
Additional information
The level unit may differ from the distance unit as defined in the Distance unit parameter
(® ä 116):
• The distance unit is used for the basic calibration ("Empty calibration" and "Full calibration").
• The level unit is used to display the (unlinearized) level.
Blocking distance
Navigation
Setup ® Advanced setup ® Interface ® Blocking dist.
Setup ® Advancede setup ® Safety settings ® Blocking dist.
Desription
Defines the upper blocking distance, UB
Input range
0 to 200 mm (0 to 7.8 in)
Factory setting
0 m (0 ft)