Description of device parameters (operating menu) Levelflex FMP51, FMP52, FMP54
134 Endress+Hauser
Additional information
Echoes from within the blocking distance are not
taken into account in the signal evaluation. The
upper blocking distance is used
•
to suppress interference echoes at the top end of
the probe (1).
• to suppress the echo of the total level in the case
of flooded bypasses (2).
A0013220
Level correction
Navigation
Setup ® Advanced setup ® Interface ® Level correction
Description
Defines the level correction
Input range Depending on the selected level unit:
• -100,0 ... 100,0 %
• -200,0 ... +200,0 m
• -656,2 ... +656,2 ft
• -7874,0 ... +7874,0 inch
• -200.000,0 ... +200.000,0 mm
Factory setting
0%
Additional information
The value specified in this parameter is added to the measured total and interface level.