Levelflex FMP51, FMP52, FMP54 Description of device parameters (operating menu)
Endress+Hauser 141
Additional information
For Table mode = Manual: Level is a writable parameter.
For Table mode = Semi-automatic: Level is a read-only parameter.
Customer value
Navigation
Setup ® Advanced setup ® Linearization ® Customer value
Condition
Only visible if the "Table" linearization type has been selected.
Description
Specification of the linearized value of the respective linearization point.
Input range
-3,0 x 10
38
... +3,0 x 10
38
Factory setting
0
Activate table
Navigation
Setup ® Advanced setup ® Linearization ® Activate table
Condition
Only visible if the "Table" linearization type has been selected.
Description
Enables or disables the linearization table.
Selection
• Disable
A linearization is not calculated. If the Linearization type parameter has been set to Table,
the error message F435 is generated.
•
Enable
The measured value is linearized according to the table before being sent to the output.
Factory setting
Disable
Additional information
When editing the table, this parameter is automatically reset to the Disable option. After finishing
the editing procedure it must be set to the Enable option again.