Levelflex FMP51, FMP52, FMP54 Mounting
Endress+Hauser 47
Step Parameter Action
3
Expert ® Sensor ® Sensor properties ® Probe length
correction ® Confirm probe length
Select Manual input option.
4
Expert ® Sensor ® Sensor properties ® Probe length
correction ® Present probe length
Enter measured probe length.
Vessels with heat insulation
If process temperatures are high, the device must be included in normal tank insulation to
prevent the electronics heating up as a result of heat radiation or convection. The insulation
may not exceed beyond the points labeled "MAX" in the drawings.
MAX
1
3
MAX
2
A0014653
å 9
Process connection with thread - FMP51
1 Tank insulation
2 Compact device
3 Sensor remote (feature 600)
40(1.57)
40(1.57)
MAX
mm(in)
MAX
1
3
2
A0014654
å 10
Process connection with flange - FMP51, FMP52
1 Tank insulation
2 Compact device
3 Sensor remote (feature 600)