Micropilot FMR60B PROFINET with Ethernet-APL System integration
Endress+Hauser 43
Module: Binary input
The Binary input module can cyclically send discrete values from the device to the
automation system. The status of the Heartbeat Verification is transmitted for the device:
Module: Binary input Heartbeat Technology slot 80
Bit Function Description
0 Status parameter Not done option Verification not carried out
1 Status parameter Failed option The device has not passed the verification. At least one test group
was outside the specifications.
2 Status parameter Busy option Verification in progress
3 Status parameterDone option Verification carried out
4 Verification result parameter Failed option The device has not passed the verification. At least one test group
is outside the specifications.
5 Verification result parameter Passed option The device has passed the verification. All of the verified test
groups corresponded to the specifications.
6 The verification result is also "Passed" if the result for an individual test group
is "Failed" and the result for all other test groups is "Passed".
7 Verification result parameter Not done option Verification not carried out
Module: Binary input Sensor diagnostics slot 81
Bit Function Description
0 Buildup on the sensor Buildup detected on the sensor
1 Foam generation Foam generation detected
2-7 -
8.3.3 Status coding
Status Coding (hex) Meaning
BAD - Maintenance alarm 0x24 No measured value available, as a device error has occurred.
BAD - Process related 0x28 No measured value available, as the process conditions are not within the
device's technical specification limits.
BAD - Function check 0x3C A function check is active (e.g. cleaning or calibration)
UNCERTAIN - Initial value 0x4F A predefined value is output until a correct measured value is available
again or corrective measures have been performed that change this
status.
UNCERTAIN - Maintenance
demanded
0x68 Wear and tear has been detected. Maintenance is needed shortly to
ensure the device remains operational.
The measured value might be invalid. The use of the measured value
depends on the application.
UNCERTAIN - Process related 0x78 The process conditions are not within the device's technical specification
limits. This could have a negative impact on the quality and accuracy of
the measured value.
The use of the measured value depends on the application.
GOOD - OK 0x80 No error has been diagnosed.
GOOD - Maintenance demanded 0xA8 The measured value is valid.
It is highly advisable to service the device in the near future.
GOOD - Function check 0xBC The measured value is valid.
The device performs an internal function check. The function check does
not have any noticeable effect on the process.