Repair Micropilot FMR60B PROFINET with Ethernet-APL
62 Endress+Hauser
13 Repair
13.1 General information
13.1.1 Repair concept
Endress+Hauser repair concept
• The devices have a modular design
• Repairs can be carried out by Endress+Hauser Service or by appropriately trained
customers
• Spare parts are grouped into logical kits with the associated replacement instructions
For more information on service and spare parts, please contact your Endress+Hauser
sales representative.
13.1.2 Repair of Ex-certified devices
L
WARNING
Incorrect repair can compromise electrical safety!
Risk of explosion!
‣
Only specialist personnel or the Endress+Hauser service team may carry out repairs on
Ex-certified devices in accordance with national regulations.
‣
Relevant standards and national regulations on hazardous areas, safety instructions
and certificates must be observed.
‣
Use only original Endress+Hauser spare parts.
‣
Please note the device designation on the nameplate. Only identical parts may be used
as replacements.
‣
Carry out repairs according to the instructions.
‣
Only the Endress+Hauser service team is permitted to modify a certified device and
convert it to another certified version.
13.2 Spare parts
• Some replaceable device components are identified by a spare part nameplate. This
contains information about the spare part.
• All the spare parts for the measuring device, along with the order code, are listed in the
Device Viewer (www.endress.com/deviceviewer) and can be ordered. If available, users
can also download the associated Installation Instructions.
Device serial number or QR code:
Located on the device and spare part nameplate.
13.3 Replacement
L
CAUTION
Data upload/download is not permitted if the device is used for safety-related
applications.
‣
After an entire device or an electronics module has been replaced, the parameters can
be downloaded to the device again via the communication interface. For this, the data
must have been uploaded to the PC beforehand using the "FieldCare/DeviceCare"
software.