Micropilot FMR60B PROFINET with Ethernet-APL Technical data
Endress+Hauser 89
Installation height as per
IEC61010-1 Ed.3
Generally up to 5 000 m (16 404 ft) above sea level
Degree of protection Test as per IEC 60529 and NEMA 250-2014
Housing
IP66/68, NEMA TYPE 4X/6P
IP68 test condition: 1.83 m under water for 24 hours.
Cable entries
• Gland M20, plastic, IP66/68 NEMA TYPE 4X/6P
• Gland M20, nickel-plated brass, IP66/68 NEMA TYPE 4X/6P
• Gland M20, 316L, IP66/68 NEMA TYPE 4X/6P
• Gland M20, hygiene, IP66/68/69 NEMA Type 4X/6P
• Thread M20, IP66/68 NEMA TYPE 4X/6P
• Thread G1/2, IP66/68 NEMA TYPE 4X/6P
If the G1/2 thread is selected, the device is delivered with an M20 thread as standard
and a G1/2 adapter is included with the delivery, along with the corresponding
documentation
• Thread NPT1/2, IP66/68 NEMA TYPE 4X/6P
• M12 plug
• When housing is closed and connecting cable is plugged in: IP66/67 NEMA TYPE 4X
• When housing is open or connecting cable is not plugged in: IP20, NEMA TYPE 1
NOTICE
M12 plug: Loss of IP protection class due to incorrect installation!
‣
The degree of protection only applies if the connecting cable used is plugged in and
screwed tight.
‣
The degree of protection only applies if the connecting cable used is specified according
to IP66/67 NEMA 4X.
‣
The protection classes are only maintained if the dummy cap is used or the cable is
connected.
Vibration resistance DIN EN 60068-2-64 / IEC 60068-2-64 for 5 to 2 000 Hz: 1.5 (m/s
2
)
2
/Hz
Electromagnetic
compatibility (EMC)
• Electromagnetic compatibility as per EN 61326 series and NAMUR recommendation
EMC (NE21)
• Maximum measured error during EMC testing: < 0.5 % of the current digital measured
value
For more details refer to the EU Declaration of Conformity.