Note:
You can also change to another
main group when you are in a
sub-group.
But then any setting you failed to
confirm by pressing E is not saved.
If no input is made in a sub-group
for longer than about 10 minutes, the
instrument automatically returns to
measuring mode (exceptions:
Calibration, Simulation and Start-up).
Hold function
In order to avoid unintentional changes to the
current outputs during configuration or
calibration, the present current value at the
current output can be “frozen” or a fixed
current can be specified by using the “Hold”
function.
Limit contactor output contacts are set to
passive by the “Hold” function
(NO contact open, NC contact closed).
The top display indicates “Hold” on the right
instead of the current output value; on the left
is the current measured value of the main
parameter as usual.
Also refer to:
• Chapter 6.7, Short Operation menu:
Hold on / off
• Chapter 7.2, Current Output menu, Hold
type:
Fixed current value / previous measured
value
• Chapter 7.4, Calibration Preset menu,
Hold at Cal.: yes / no
PA RA M
MEAS CAL
DIAG
Start main group
by pressing
appropriate key
Select subgroup
and press E
increment
Select option:
decrement
Select digit
Parameter 4
Parameter 3
Last parameter
of subgroup
Parameter 2
Subgroup
Subgroup
Subgroup
Main group
Option
Option
Option
Parameter 1
Select
option
Set numeric value
123.45
Parameter 1
PARAM
PARAM
PARAM
PARAM
PARAM
BEDSCHEE.CDR
Fig. 6.3
Block diagram of
Mycom-Operating concept
lm152e06.chp
Mycom CLM 152 Operation