7.4 Calibration presettings
7.5 Clean function
The »Chemoclean« clean function requires a
device with a total of five output contacts.
The CLA 111 assembly and the CLR 30 spray
head are required to automate the cleaning
functions.
Contact 3 (terminal 91/92) is used for the
motive water, contact 4 (terminal 93/94) is for
dosing of cleaner.
The »Chemoclean« clean function must be
set in the »System configuration / Output
contacts« menu (see Chapter 7.1.4).
→ Instrument data
→ Calibration
Function Selection Factory setting
Hold for calibration yes, no yes
Enter temperature coefficient Tc of
calibration solution
0.00 ... 10.00 %/K 2.10 %/K
Enter temperature of calibration
solution (only for MTC)
–35.0 ... +250.0 °C25.0°C
→ Instrument data
→ Chemoclean
PARAM
PARAM
Instrument configuration Mycom CLM 152
50 Endress+Hauser