11.3 Error table
Failure
No. Display Corrective action
E001 Fault in internal communication Send instrument to Endress+Hauser
agency responsible for repair or call
Service.
If error »07« occurs at the same time,
check transmitter acc. to this error.
E002 Data error in EEPROM
E003 Invalid configuration Check configuration of slots
E004 Changed configuration
Accept new configuration using Set config
in Service / Factory settings menu
E005 Unknown card code Check configuration of slots
E006 Checksum error in EEPROM
Carry out checksum correction in Special
functions
E007
Transmitter faulty or transmitter function
disturbed
Check sensor connection
Check if measuring range is exceeded
Check on exceeded sensor cross-talk
Check temperature measurement. If
temperature measurement is not in
order, replace the module
E010 Temperature sensor defective Check temperature measurement and
terminals; poss. check measuring
transmitter and measuring cable with
temperature simulator
E011 Temperature sensor 2 defective
E015 Current loop 1 open
Check terminals, lines and poss.
connected devices
E016 Current loop 2 open
E017 Sensor error
(inductive only)
Send sensor to your Endress+Hauser
agency for examinaton or request Service
E018 Sensor error channel 2
(inductive only)
Maintenance required
E036 Calibrating range of sensor 1 exceeded
Recalibrate sensor; poss. check sensor
and terminals; check measuring
transmitter and measuring cable with CD
simulator
E037 Calibrating range of sensor 1 undershot
E038 Calibrating range of sensor 2 exceeded
E039 Calibrating range of sensor 2 undershot
E040 TC table invalid
Correct temperature compensation table;
check temperature measurement and
terminals; poss. check measuring
transmitter and measuring cable with
temperature simulator
E041 Substance table invalid correct substance table
E049
Adaptation factor range of sensor 1
exceeded
Enlarge the distance of sensor from wall
E050
Adaptation factor range of sensor 1
undershot
Enlarge the distance of sensor from wall
E051
Adaptation factor range of sensor 2
exceeded
Enlarge the distance of sensor from wall
E052
Adaptation factor range of sensor 2
undershot
Enlarge the distance of sensor from wall
Instrument diagnostics Mycom CLM 152
70 Endress+Hauser

Table of Contents

Other manuals for Endress+Hauser Mycom CLM 152

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Endress+Hauser Mycom CLM 152 and is the answer not in the manual?

Endress+Hauser Mycom CLM 152 Specifications

General IconGeneral
BrandEndress+Hauser
ModelMycom CLM 152
CategoryTransmitter
LanguageEnglish

Related product manuals