8.1 Instruments with two contacts
Cell dependency
Instrument data
Limit contactor
Function Selection Factory setting
Select group
Limit configuration
Alarm configuration
Operating mode
Limit
configuration
If Limit configuration selected
Switch on/off output on, off off
Limit see Table Cell dependence
Hysteresis see Table Cell dependence
Contact function
Min function
Max function
Min function
On delay 0 ... 7200 s 0 s
Off delay 0 ... 7200 s 0 s
Contact type NC contact, NO contact NO contact
If Alarm configuration selected
Alarm threshold see Table Cell dependence
Alarm delay 0 ... 6000 s 0 s
If Operating mode selected
Switch over operating mode
Auto limit contactor 1
Manual limit contactor 1
Manual
limit contactor 1
Limit contactor manual mode
(if Manual selected)
off, on off
Limit Hysteresis Alarm
Selection Default Selection Default Selection Default
CLS 50 0
µ
S/cm 1000 mS/cm
Limit 1: 50.0 mS/cm
Limit 2: 950 mS/cm
0.1
µ
S/cm 200.0 mS/cm 20.00 mS/cm 0.1
µ
S/cm 1000 mS/cm 50.0 mS/cm
CLS 52 0
µ
S/cm 1000 mS/cm
Limit 1: 50.0 mS/cm
Limit 2: 950 mS/cm
1
µ
S/cm 200.0 mS/cm 20.00 mS/cm 1
µ
S/cm 1000 mS/cm 50.0 mS/cm
k = 0.01 0 nS/cm 200.0
µ
S/cm
Limit 1: 10.00
µ
S/cm
Limit 2: 190.0
µ
S/cm
0.1 nS/cm 20.00
µ
S/cm 2.000
µ
S/cm 0.1 nS/cm 200.0
µ
S/cm 10.0
µ
S/cm
k = 0.1 0
µ
S/cm 2000
µ
S/cm
Limit 1: 100.0
µ
S/cm
Limit 2: 1900
µ
S/cm
0.001
µ
S/cm 200.0
µ
S/cm 20.00
µ
S/cm 0.001
µ
S/cm 2000
µ
S/cm 100.0
µ
S/cm
k=1 0
µ
S/cm 20.00 mS/cm
Limit 1: 1000
µ
S/cm
Limit 2: 19.00 mS/cm
0.01
µ
S/cm 2000
µ
S/cm 200.0
µ
S/cm 0.01
µ
S/cm 20.00 mS/cm 1000
µ
S/cm
PARAM
Limit configuration Mycom CLM 152
54 Endress+Hauser