11.7 Service
11.7.1 Simulation
Note:
In the „Simulation“ menu the displayed
value for current output or the
displayed relay state is immediately
displayed. If you change the value in
the window, the current output or the
relay state changes at the same
time. If you exit the window, the
simulation is deactivated and the
current outputs and relay states
are reset to the current values.
11.7.2 Internal data
→ Service
→ Simulation ☞ 11.7.1
→ Internal data ☞ 11.7.2
→ Factory settings ☞ 11.7.3
→ Instrument check ☞ 11.7.4
→ Special functions ☞ 11.7.5
Function Selection
Set present
current value at current output 1
0.00...22.00 mA
Set present
current value at current output 2
0.00...22.00 mA
Set current relay states
(The number of contacts displayed is
dependent on the instrument expansion level
and configuration)
Select contact by pressing ↑ and ↓.
Open / close selected contact by pressing →.
Measured value simulation
Main measured value
Free setting across configured measuring
range (sensor independence)
Change current outputs and relay states
depending on their configuration
Measured value simulation
Temperature
–35.0 ... +250.0 °C
The current output changes depending on its
configuration
Function Selection
Display instrument number No selection
Display software version of instrument No selection
Display software version of FCL1 module
(CD signal processor) channel 1
No selection
Display software version of FCL1 module
channel 2 (for difference)
No selection
Display hardware configuration in several
consecutive windows: module, installation
date, slot assignment
No selection
Order Code
Alphanumeric entry with characters 0 ... 9 and
a...Z
Reset counter 0 ... 255 display only
DIAG
Instrument diagnostics Mycom CLM 152
74 Endress+Hauser