7.1 System configuration
Instrument data
System configuration
Measured variable 7.1.1
Code 7.1.3
Output contacts 7.1.4
General settings 7.1.5
7.1.1 Measured variable
PARAM
Function Selection Factory setting
Select measuring type
Conductivity, concentration, not
selectable with differentiation),
resistance
Conductivity
Select sensor
CLS 50, CLS 52,
2-ring:k=0,01;k=0,1;k=1;k=10
CLS 52 for CD, conc.
2-ring k = 0,01 for
resistance
Differential or single-circuit
measurement
(only on instruments with two
measuring channels)
Differential measurement,
Single-circuit measurement
Differential
measurement
Select unit
(mS / cm), (mS / cm) / (mS / m), (S / m)
(k cm), (M cm) / ( m), (k m)
(µS / cm), (mS / cm)
for CD, conc.
(k cm), (M cm)
for MOhm
Select temperature sensor 1
PT 100
PT 1000
NTC 30 k
PT 100
Select temperature sensor 2
(only for difference)
PT 100
PT 1000
NTC 30 k
PT 100
Configure input
contacts (only for FCXI expansion
module, remote measuring range
switch-over see 7.1.2)
2 x hold,
1 x hold / 1 x remote-switch-over
2 x remote-switch-over
external Clean control
*)
2 x hold
*)
Only in relay configuration with Clean function!
lm152e07.chp
Mycom CLM 152 Instrument configuration
Endress+Hauser 35

Table of Contents

Other manuals for Endress+Hauser Mycom CLM 152

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Endress+Hauser Mycom CLM 152 and is the answer not in the manual?

Endress+Hauser Mycom CLM 152 Specifications

General IconGeneral
BrandEndress+Hauser
ModelMycom CLM 152
CategoryTransmitter
LanguageEnglish

Related product manuals