7.1 System configuration
→ Instrument data
→ System configuration
→ Measured variable ☞ 7.1.1
→ Code ☞ 7.1.3
→ Output contacts ☞ 7.1.4
→ General settings ☞ 7.1.5
7.1.1 Measured variable
PARAM
Function Selection Factory setting
Select measuring type
Conductivity, concentration, not
selectable with differentiation),
resistance
Conductivity
Select sensor
CLS 50, CLS 52,
2-ring:k=0,01;k=0,1;k=1;k=10
CLS 52 for CD, conc.
2-ring k = 0,01 for
resistance
Differential or single-circuit
measurement
(only on instruments with two
measuring channels)
Differential measurement,
Single-circuit measurement
Differential
measurement
Select unit
(mS / cm), (mS / cm) / (mS / m), (S / m)
(kΩ cm), (MΩ cm) / (Ω m), (kΩ m)
(µS / cm), (mS / cm)
for CD, conc.
(kΩ cm), (MΩ cm)
for MOhm
Select temperature sensor 1
PT 100
PT 1000
NTC 30 kΩ
PT 100
Select temperature sensor 2
(only for difference)
PT 100
PT 1000
NTC 30 kΩ
PT 100
Configure input
contacts (only for FCXI expansion
module, remote measuring range
switch-over see 7.1.2)
2 x hold,
1 x hold / 1 x remote-switch-over
2 x remote-switch-over
external Clean control
*)
2 x hold
*)
Only in relay configuration with Clean function!
lm152e07.chp
Mycom CLM 152 Instrument configuration
Endress+Hauser 35