Function Selection Factory setting
Switch on/off clean function, set
parameters
Automatic ON
Automatic OFF
Settings
Automatic OFF
If „Setting» selected
Type of cleaning program Interval cleaning, Week program
Interval
cleaning
If »Interval cleaning« selected
Cleaning cycle 0.1 ... 99 h (resolution 0.1 h / 6 min)
8h
If »Week program« selected
Parameter week program Set, Cancel
Set
Week program
Number of starts (0 ... 12)
individually for every weekday
(Mon ... Sun)
0
Day program
Start time individually for every start
number
00 h 00 min
Next day Edit next weekday, Cancel
Pre-rinse time 0 ... 999 s
10 s
Cleaning time 0 ... 999 s
5s
Post-rinse time 0 ... 999 s
10 s
Repeat rate 0 ... 5 times
0
Number of cleaning cycles without
cleaning agent
0 ... 9 times
0
Hold for Chemoclean ON, OFF
ON
Hold continuation time
(measurement settling time)
0...999s
10 s
Total cleaning time no selection
00 h 00 m 00 s
lm152e07.chp
Mycom CLM 152 Instrument configuration
Endress+Hauser 51