The USP function can be activated from
the Instrument data/USP menu.
After activation, you can make the following
settings:
Function Selection Factory setting
On delay 0 ... 60 s 0 s
Hysteresis
Setting range sensor-dependent
(see Table)
0 µS/cm
Cell MR 1
1)
MR 2
1)
MR 3
1)
MR 4
1)
k = 0.01 0.0 ... 200.0 nS/cm 0.200 ... 2.000 µS/cm 2.00 ... 20.00 µS/cm 20.0 ... 200.0 µS/cm
k = 0.1 0.000 ... 2.000 µS/cm 2.00 ... 20.00 µS/cm 20.0 ... 200.0 µS/cm 200 ... 2000 µS/cm
k = 1 0.00 ... 20.00 µS/cm 20.0 ... 200.0 µS/cm 200 ... 2000 µS/cm 2.00 ... 20.00 mS/cm
k = 10 0.0 ... 200.0 µS/cm 200 ... 2000.0 µS/cm 2.00 ... 20.00 mS/cm 20.0 ... 200.0 mS/cm
1)
The selection of measuring range is executed automatically with internal measuring steps to ensure optimum accuracy
and resolution.
lm152e08.chp
Mycom CLM 152 Limit configuration
Endress+Hauser 57