Warnings
No. Display Corrective action
E080 Range for current output 1 too small
Enlarge range in Current outputs
E081 Range for current output 2 too small
E142 Knee point outside current output range 1
Correct cnfiguration in Current outputs
E143 Knee point outside current output range 2
E144
Current output range 1 too small for
selected MR (= measuring range)
Make range in Current outputs menu
larger
E145
Current output range 2 too small for
selected MR (= measuring range)
E148
Knee point outside current output
range 1 (MBX = current measuring range)
Correct configuration in Current outputs
menu
E149
Knee point outside current output
range 2 (MBX = current measuring range)
E150 Measured value outside USP table
E151 Temperature outside limits for USP table
Instrument diagnostics Mycom CLM 152