Proline Promag P 100
Endress+Hauser 19
Connection version Modbus RS485, for use in non-hazardous areas and Zone 2/Div. 2
Order code for "Output", option M
Depending on the housing version, the transmitters can be ordered with terminals or device plugs.
Order code for
"Housing"
Connection methods available
Possible options for order code
"Electrical connection"
Output
Power
supply
Option A Terminals Terminals • Option A: coupling M20x1
• Option B: thread M20x1
• Option C: thread G ½"
• Option D: thread NPT ½"
Option A Device plug
(→ 21)
Terminals • Option L: plug M12x1 + thread NPT ½"
• Option N: plug M12x1 + coupling M20
• Option P: plug M12x1 + thread G ½"
• Option U: plug M12x1 + thread M20
Option A Device plug
(→ 21)
Device plug
(→ 21)
Option Q: 2 x plug M12x1
Order code for "Housing":
Option A: compact, coated aluminum
A0019528
4 Modbus RS485 terminal assignment
1 Power supply: DC 24 V
2 Modbus RS485
Order code for
"Output"
Terminal number
Power supply Output
2 (L-) 1 (L+) 27 (B) 26 (A)
Option M DC 24 V Modbus RS485
Order code for "Output":
Option M: Modbus RS485, for use in non-hazardous areas and Zone 2/Div. 2